Top Brace and Reference Bracket

Removal

1. Remove the instrument outer case. Refer to the
2. Refer to
3. To remove the reference bracket (2), use the T-10 driver to remove the six
Figure 13-6
Top Brace and Reference Bracket Removal
5
6 places

Replacement

1. Refer to
Chapter 13
procedure.
Figure
13-6. To remove the top brace (1), use the T-10 driver to
remove the eight screws (3) (0515-0372) attaching the top brace to the
chassis, the six screws (4) (0515-1227) attaching the top brace to the boards,
and the four screws (4) attaching the top brace to the reference bracket.
screws (5) (including two at the rear panel) attaching the reference bracket to
the instrument and the four screws (4) attaching the top brace to the reference
bracket.
4
2
4
1
Figure
13-6. To replace the top brace or the reference bracket, place
them into the correct position and attach the appropriate screws. Torque to 9
inch-pounds.
Assembly Replacement Procedures
Top Brace and Reference Bracket
Instrument Outer Case
4
removal
3
8 places