Assembly Replacement Procedures

Top Brace

Top Brace

Removal

1. Remove the instrument outer case. Refer to the
2. Refer to
Figure 15-6
Top Brace Removal

Replacement

1. Refer to
450
removal procedure.
Figure
15-6. To remove the Top Brace, use the T-10 driver to
remove the eight screws (3) (0515-0372) attaching the Top Brace to the
chassis. Remove the twelve flathead screws (4) (0515-1946) attaching
the Top Brace to the chassis and boards.
Figure
15-6. To replace the Top Brace place it into the correct
position and attach the appropriate screws. Torque to 9 inch-pounds.
"Instrument Outer Case"
Keysight N9010B EXA Signal Analyzer Service Guide