Tests and Adjustments
Adjustments
Procedure
1. Press UTILITY
click EE Default Adjustment.
2. Follow the instructions and prompts as they are displayed.
IF Response Adjustment (Available Only with Option S93093/4A/B
and or S93900/1/2/4/5/7A/B Installed)
In A models these options were 090 and 093 or 094.
The IF Response adjustment is required when upgrading to Option 090, 093
and or 094. In addition, this adjustment must be performed each time a new
configuration is set up or if cables are changed:
Procedure
1. Press UTILITY
click IF Response Adjustment.
2. Follow the instructions and prompts as they are displayed.

LFE Receiver Adjustment (Option 425)

The LFE Receiver adjustment is required when upgrading to Option 205, 220,
405, 420, and 425. In addition, this adjustment must be performed each time a
new configuration is set up or if cables are changed.
Procedure
1. Press UTILITY
click LFE Receiver Adjustment.
2. Follow the instructions and prompts as they are displayed.
1. Your model PNA may not have available all of the LFE Options listed.
Keysight N5247B Service Guide
Service
System
, then
Service
System
, then
1
Service
System
, then
Adjustment Routines
, then
Adjustment Routines
, then
Adjustment Routines
, then
, then
, then
, then