Tests and Adjustments
Adjustments
Procedure
1. Press UTILITY
Default Adjustment.
2. Follow the instructions and prompts as they are displayed.

IF Response Adjustment (Option 090, 093, or 094 Only)

The IF Response adjustment is required when upgrading to Option 090, 093
and or 094. In addition, this adjustment must be performed each time a new
configuration is set up or if cables are changed:
Procedure
1. Press UTILITY
Response Adjustment.
2. Follow the instructions and prompts as they are displayed.
Keysight N5227A Service Guide
Service
System
, then
Service
, then
System
Adjustments
, then
, then click EE
Adjustments
, then
, then click IF
3-47