Configuring Measurement Conditions (Display and Function Related Settings)
CORRECTION page
Description of Softkeys
To enable/disable or otherwise control the behavior of correction based on
specified frequency points, use the following softkeys:
Softkey
ON
OFF
MEAS OPEN
MEAS SHORT
MEAS LOAD
Relationships between Correction Based on All Frequency Points and Correction
Based on Specified Frequency Points
Figure 4-13
correction features are related to each other.
Figure 4-13
How correction features are related to each other
1. This area is used to:
2. This area is used to:
128
Description
Enables the current measurement point number.
Disables the current measurement point number.
Starts open correction.
Starts short correction.
Starts load correction.
shows the CORRECTION page and how the open/short/load
Carry out correction calculation using either the open/short correction
data generated by the interpolation method or the open/short/load
correction data at specified frequency points.
Which data is used depends on the test frequencies.
Obtain the open/short correction data.
Obtain the open/short/load correction data at each frequency point
specified in the FREQ field.