Preparation for Accurate Measurement (Executing Correction)
Executing OPEN/SHORT/LOAD Correction
Reading/writing correction data (saving/recalling correction conditions)
After saving the correction conditions to a file, you can recall these conditions
on the E4981A at any time from the file.
The correction data is set up in the parameter-oriented format shown in the
table below:
Type of correction
OPEN correction
SHORT correction
LOAD correction
To read/write the correction data, use the following command.
•
The correction data written using the above command is handled as follows:
•
•
•
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Parameter format
G-B or Cp-G
(select with the
[:SENSe]:CORRection:CKIT:STAN1:FORMat
command)
R-X or Ls-Rs
(select with the
[:SENSe]:CORRection:CKIT:STAN2:FORMat
command)
Parameter format to define the LOAD correction standard
(select with the
[:SENSe]:CORRection:CKIT:STAN3:FORMat
command)
[:SENSe]:CORRection:DATA
In case of STAN3: Saved as LOAD correction data for the measurement
frequency specified when the command is executed.
In case of STAN1/STAN2: Saved as OPEN/SHORT correction data for all
frequencies.
Saved as the data for normal operation when the multi-correction function
is OFF and as the data for the multi-correction when it is ON (for the
channel that has been selected at execution).
on page 240
51