Acquisition Parameters; Acquisition Band (With 81Dispap Modules); Averaging Acquisition; Power Level - Viavi 8100 V2 Series User Manual

Portable, modular platform designed for the construction, validation and maintenance of optical fiber networks
Hide thumbs Also See for 8100 V2 Series:
Table of Contents

Advertisement

Chapter 13 Attenuation profile
AP Activation and self calibration

Acquisition parameters

Acquisition band (with 81DISPAP modules)

OESCL
SCL
NOTE
In case of OBS5xx handheld sources, the measurement will be performed
exclusively on SCL band.

Averaging acquisition

It enables to improve the dynamic range of the measurement by reducing the noise
level. It is recommended to use the Auto mode and configure a manual averaging if
needed only:
No
Low
Medium
High

Power Level

This parameter enables to get a power level indication at the start of an acquisition (only
with 81DISPAP or 81MRDISPAP modules)
No
Yes
NOTE
If Yes is selected, the acquisition stops if not enough power is received.
410
measurement performed over the full wavelength range (OESCL
bands).
measurement performed over a wavelength range limited to S,C and
L bands
No average of the acquisition sample to be performed.
Low averaging (4 samples).
Medium averaging (16 samples).
High averaging (32 samples).
the received power is not indicated before starting the measurement
acquisition.
the received power is indicated at the beginning of measurement.
User Manual
78000010220 Rev. 024

Advertisement

Table of Contents
loading

Table of Contents