Each test equipment must set its own references and conform to the following process:
1
Press R
2
Press References > Take Refs keys and follow the step by step instructions to
perform references on each unit.
3
Choose between side by side or loopback for you loss referencing method.
Figure 124
Select the reference to be performed
Loopback Referencing method
The loopback referencing is used when the two units are at different location.
After clicking on Loopback, the wizard will guide you through two steps:
1
The self reference is used for loss and ORL testing. Connect the jumper from the
module port to the mainframe powermeter and press Ok to start referencing.
button
ESULTS
User Manual
Chapter 7 FiberComplete Modules
78000010220 Rev. 024
Establishing References