Acquisition Parameters - Viavi 8100 V2 Series User Manual

Portable, modular platform designed for the construction, validation and maintenance of optical fiber networks
Hide thumbs Also See for 8100 V2 Series:
Table of Contents

Advertisement

Chapter 11 I-PMD Measurement
I-PMD Activation and Configuration

Acquisition parameters

In the Setup screen,
1
First, select PMD/I-OSNR on the parameter Mode,
2
Then select the following parameters for acquisition:
PMD Mode
Select the acquisition mode in the parameter:
Channel
Grid
BBS
Sweep
HR OSA Acquisition will be performed on a given frequency range (see sweep range
menu to a complete description). In this mode, OSNR are measured «out band».
Channel Frequency
If Channel is selected in the Mode parameter, adjust the frequency of the channel to
analyze (using the Edit Number menu key).
The Wavelength parameter is automatically configured according to channel
frequency defined.
Sweep range
If Sweep mode is selected in the Mode parameter, select the frequency range in this
sub-menu:
Full
Start/End
Center/Span
360
Acquisition will be performed on one G.694.1 Channel
Acquisition will be performed on multiple channels defined in a grid
Acquisition will be performed with BBS inserted, instead of Transmis-
sion signals (intrusive measurement, network should be down and
replaced by a BBS module - see
BBS").
Full frequency range is used for measurement
Select manually the Start/End frequency
Select manually the Center/Span frequency