On/Off-OSNR Method
Challenge
This is a precise out-of-service method used for measuring the in-band OSNR of stan-
dard and also of polarization-division multiplexed (PDM) systems.
Principle of "On/Off-OSNR Measurement"
The On/Off-OSNR method is based on measuring the noise power when the transmis-
sion channel is switched-off.
This is a 2 step method:
1
Switch-on all channels and perform a standard WDM measurement
All parameters like power (P
instrument
2
Switch-off the channel carrying PDM signals and perform a second measure-
ment
The channel power measured at the deactivated channel wavelength will indicate
the in-band noise P
In-band OSNR is calculated based on P
Setup
Select 'Measurements Mode' = OO-OSNR'
Figure 147
OSA Setup: OO-OSNR configuration
Set Ref. Channel
on
= noise power
off