Setup Menu - Viavi 8100 V2 Series User Manual

Portable, modular platform designed for the construction, validation and maintenance of optical fiber networks
Hide thumbs Also See for 8100 V2 Series:
Table of Contents

Advertisement

1
Press S
ETUP
2
On bottom right of setup page, press Load Config. menu key.
3
In the Explorer, select the desired file configuration
4
Press
Load
A beep is emitted to validate the selection of the configuration file.
The software automatically brings you back to setup page.
NOTE
Most of the configuration files are available into the Platform in disk/con-
fig/PMD.

Setup menu

To access the PMD test setup menu, press the S
You can choose the default values by pressing the Test Auto key or define your own
configuration.
The softkey is available when one header is selected.
In Test Auto configuration, the setups below are provided.
Acquisition
1
Sweep
: Single
Averaging acquisition: Auto
Long term: No
Analysis
Coupling: Strong
Amplified Link: No
Measurement Band; Auto
Link
Fiber Nbr Increment: Yes
1.Does not apply with 81DISPAP/81MRDISPAP modules used with a PSM
Chapter 10 Polarization Mode Dispersion Measurement
hard key.
Config. menu keys.
User Manual
PMD Activation and self calibration
button on the Platform 8000.
ETUP
78000010220 Rev. 024
335