Viavi 8100 V2 Series User Manual page 545

Portable, modular platform designed for the construction, validation and maintenance of optical fiber networks
Hide thumbs Also See for 8100 V2 Series:
Table of Contents

Advertisement

1
On the R
2
Push the
3
Link the FOX connectors of the two MTS 8000s by using two jumpers.
4
Push on
Figure 259
Taking a side by side reference (8000 V2 series example)
Reference date.
Type & serial number
of the distant
instrument
Once the references are taken, the reference screen is updated for the wave-
lengths available on the distant instrument. It indicates the type of reference
used.
The date of the reference as well as the type and serial number of the distant
instrument are indicated on the first line of the table.
If the reference is correct, disconnect the jumpers at the middle point, in order to insert
the fiber to be tested.
If the reference taken is not correct, take a new reference.
If the reference measurements are too weak, a warning is displayed.
Clean all connections to improve the reference measurements.
The reference is only valid if the jumpers are not disconnected from the MTS
8000.
page, push the
ESULTS
Loss Side/Side
key.
Go
key.
User Manual
References
key.
78000010220 Rev. 024
Chapter 18 OFI Module
FOX Function
509