Chapter 7 FiberComplete Modules
Establishing References
Figure 125
Self reference
Powermeter of the
8000 V2 platform
The reference values are stored and displayed at the end of referencing.
2
The zero ORL reference is necessary for ORL testing. Once the self reference
measurement has been carried out, the Zero ORL adjustment can be performed
Connect the jumper from the module port to the non-reflective termination via a
mating sleeve. If you don't have a non-reflective termination, a mandrel can be
used. Press Ok to start referencing.
222
User Manual
OTDR Module port
78000010220 Rev. 024