Chapter 12 High Resolution OSA Measurement
Overlay trace function
Changing the traces position
Once a trace is displayed in overlay, the traces can be adjusted according to the Y axis:
1
Press Advanced > Overlay menu keys.
2
Select the adjustment according to Y axis:
Y
Reset: all traces are on the the same level at the intersection with the active
cursor.
Y Shift: Each trace is shifted from 5 dB from the other.
Y Exact: the traces displayed are on the same position according to their injec-
tion level.
Trace resulting from the difference between two
traces
It is possible to obtain the trace corresponding to the point-by-point difference between
the current trace and the trace in overlay (if only two traces are displayed simultane-
ously).
To do this:
1
Press Advanced > Overlay menu keys
2
Press the
The screen displays the two traces in overlay and the trace resulting from the
"Difference".
Removing a trace in overlay
Removing the current trace
It is possible to remove a trace displayed. To do this:
1
Select it (see previous paragraph),
2
In Results page, press Advanced > Overlay menu keys
3
Press
398
2 Curves Diff.
Remove Current
Trace.
User Manual
menu key.