Here is a test setup for in-band OSNR testing.
Figure 167
Test setup for In-band OSNR testing
Example:
•
OSNR measured with OSCA-710 at terminal site (1) provides value of 14dB
•
Service failed as the OSNR is < 20 dB
How to locate the failure?
Perform the following tests at EDFA monitor access points (2-5):
•
Check per channel input and output power of EDFA
=> Input power must be in the system specified range
•
Check power uniformity
=> Equal power levels for all channels at EDFA output
•
Compare OSNR from EDFA to EDFA
=> OSNR may decrease because of the amplifier noise figure NF (type 3-4 dB)
per EDFA
•
Locate and exchange optical amplifier