Chapter 11 I-PMD Measurement

Performing a Zero Reference

1
If necessary, press S
2
Select one parameter in one of the setup page (acquisition, link..)
3
Press Save Config. menu key
4
Enter a name for the configuration file using the edition keypad (max 20 charac-
ters).
NOTE
Configuration file is saved in the directory disk/config/IPMD.
5
Press Enter to validate
A sound is emitted to indicate the file is saved.
The configuration file is saved with the extension ".fo_cfg" (icon
recalled at any time from the Explorer page.
Performing a Zero Reference
For High OSNR accuracy, it is recommended to perform a Zero Referencing before first
use of the product:
1
Press S
2
Put caps on the module optical port
3
Press
4
Confirm the reference process pressing Yes.
364
ETUP
hard key to reach Setup screen
ETUP
Zero Ref.
menu key
User Manual
to return to Setup page.