Il/Orl Referencing; Il/Orl Referencing With The Integrated Fibercomplete® Solution; Orl And Side By Side Il Referencing - Viavi 8100 V2 Series User Manual

Portable, modular platform designed for the construction, validation and maintenance of optical fiber networks
Hide thumbs Also See for 8100 V2 Series:
Table of Contents

Advertisement

IL/ORL referencing

IL/ORL referencing with the integrated
FiberComplete® solution
1
Connect a fiber jumper to the MTAU COM port of T-BERD/MTS 8000 V2 and one
to the test port of the module of the T-BERD/MTS 6000A.
Refer to
Figure 229 on page 461
2
Press O
/O
N
Activate the FCOMP function on both units
1
Press the H
2
Press the hard key R
3
Press the soft key

ORL and side by side IL Referencing

1
Touch the FCOMP tab.
2
Press the soft key References to enter the reference menu.
3
Press the soft key Take Refs in order to enter the referencing wizard.
4
Press Side/Side key to start referencing process for ORL and side by side Inser-
tion Loss.
5
When prompt, connect the test jumpers from the 8000V2 COM port to the main-
frame power meter and from the 6000 test module to mainframe power meter, as
shown below:
to turn both units on and wait the completion of auto-test (45 sec.).
FF
hard key and light the icon
OME
and touch the MTAU tab of the 8000 V2 unit.
ESULTS
until OUT A is selected on the 8000 V2 unit.
User Manual
Chapter 17 Fiber Characterization
Referencing the test functions
for proper module interconnections.
78000010220 Rev. 024
.
463

Advertisement

Table of Contents
loading

Table of Contents