Chapter 17 Fiber Characterization
Fiber Characterization Automated Test Sequence using the Link Characterization Script
1
Press the S
2
Go to the FCOMP tab on both units
3
Configure the FCOMP Thresholds parameters as Default or enter your own
criteria
4
Go to the SM-OTDR tab on both units
5
Press the soft key Alarms
6
Select Alarm level to "Fail"
7
Set the thresholds to Default or define your own criteria
Figure 243
OTDR Alarms - Default and User
Selecting data storage directory
1
Press the F
2
Press File/Explorer softkey to select the "Explorer" page.
3
Using the navigation keys or touchscreen, on the left side of the screen, select
the main or sub directory where you wish to store files, typically the "Hard Disk".
4
If necessary, create a new directory to store the results: see
directory" on page
Launching the Link Characterization script
1
Press the S
2
Select the function Scripts > Link Characterization using touchscreen or stylus
3
Press the soft key Launch to enter the script configuration menu.
484
hard key on both units
ETUP
hardkey.
ILE
474.
hard key on the 8000 V2 and 6000A units
CRIPT
User Manual
78000010220 Rev. 024
"Creating the storage