If a CD mode is selected, the key Full Band/SCL Band is displayed and allow to
select transmitted wavelength range.
Configuration with the 81BB1A module
Figure 223
Configuration of the BBS1A
1
Select CD, AP/PMD or ALL transmission mode to perform the corresponding test
using the 81DISPAP or 81MRDISPAP module.
If the parameter All is selected, the dynamic range will be degraded by
3 dB.

Remote interlock connector

The use of a Remote Interlock system is specified with class IIIb sources by 21 CFR
1040.10 (USA).
The BBS module is equipped with a remote interlock connector (SMB type) on the front
panel. This is aimed to protect the user from injury when using class IIIb light sources.
If the short circuit at the SMB connector is opened, the broadband light source is
switched off immediately and cannot be switched on until it is closed again.