The following parameters, defined on master unit, are automatically
applied/transferred to the slave unit: Laser - IL/ORL Bidir. - OTDR
Acquisition Auto (if OTDR Acquisition is defined on Manual, both units
will performed a manual measurement, but according to their own configura-
tion in the OTDR tab: the parameters can then be different).

Alarms parameters

In the Setup page, press Alarms softkey (if one parameter is selected in the current
screen, press Top Menu soft key to display the right menu keys and click on Alarms).
Thresholds
-
-
-
-

Analysis parameters

In the Setup page, press Analysis softkey (if one parameter is selected in the current
screen, press Top Menu soft key to display the right menu keys and click on Analysis).
Unit
Select the unit for the distance measurement: km / kfeet / miles.
Table View
Allows to choose the kind of table to be displayed at the end of acquisition.
Fiber
Select the user defined thresholds to be used: User 1/ User 2 /
User 3 / User 4 and enter limits for:
Loss: enter a loss threshold for each wavelength (dB)
ORL: enter an ORL threshold for each wavelength (dB)
Or select the Default parameter to define thresholds by default for
Loss and ORL values:
Loss: > 40 dB for each wavelength
ORL: < 27 dB for each wavelength
Select None if alarm thresholds must not be defined
displays detailed results for one fiber: Loss B -> A and loss A -
> B at each wavelength; the average loss at each wavelength
and the ORL A and ORL B at each wavelength. See
View" on page