Acquisition Parameters - Viavi 8100 V2 Series User Manual

Portable, modular platform designed for the construction, validation and maintenance of optical fiber networks
Hide thumbs Also See for 8100 V2 Series:
Table of Contents

Advertisement

NOTE
The selection of the RDZ function automatically de-selects the selection of
the Multimode function on the same module ,and vice-versa.

Acquisition parameters

Acquisition
Range
Select the range of the measurement: from 50 m to 1km
or
Click on Manual to enter manually a range.
or
Click on Auto to detect automatically the range.
In Auto mode, the range is selected as a function of the end of the fiber.
Time
Select one of the acquisition times predefined: 10 seconds / 20 seconds / 30 seconds /
1 minute.
Otdr connector test
This parameter allows to choose if a test of the front connector must be performed when
acquisition is launched.
No
the OTDR connection is tested with indication Bad/Good.
Yes & Continue
the OTDR connection is tested, and if the state is not good, the
acquisition continues but a warning displays.
Yes & Abort
the OTDR connection is tested, and if the state is bad, a warning
displays and the acquisition stops.
Launch cable
Launch Cable End / Receive Cable Start
No
All the results are displayed and referenced on the basis of the board
of the module.
Chapter 4 Reduced Dead Zone OTDR application
User Manual
Selecting and configuring the RDZ function
78000010220 Rev. 024
169