Fiber Characterization Automated Test Sequence using the Link Characterization Script
Starting a test sequence
1
Press the soft key Start on both units
A message Waiting for Datalink initialisation... is displayed
while both units are communicating to each others in order to synchronize the test
sequence.
2
Inspect and clean connectors of the fiber to test using the P5000i inspection
scope on both ends when the inspection page appears.
3
Connect the fiber under test to each patchcord (one connected to the TB8000
COM port of the MTAU module and one connected the 6000A module C).
4
Press the S
CRIPT
On the T-BERD/MTS 8000 V2
1
The message
between both test sets is displayed.
NOTE
The test set will start the test sequence automatically. Do not press the Yes /
No keys.
On the T-BERD/MTS 6000A V2
1
The following message appears while the 8000 V2 test is in progress:
DO NOT click Stop until the 8000 V2 notifies test is completed.
On the T-BERD/MTS 8000 V2
1
The message
hard key on both units to continue with the test sequence.
User Manual
Chapter 17 Fiber Characterization
displays until the green link
to start PMD/CD/AP measurements display.