With the OptiPulses option, the screen is as follows:
–
Two traces for each wavelength measured: the trace with shortest pulse and
the standard one,
–
One single results table per wavelength with results from both traces merged.
Example: if OTDR acquisition if performed on two wavelengths, the screen
displays 4 traces and 2 results table.
Click on one event in the results table to automatically turns the trace on which it has
been detected active
OptiPulses and SmartLink option
The SmartLink screen displays the graphical representation of the link with all the events
detected, whether they are detected on the trace with shortest pulse or on standard
trace.
Figure 68
OptiPulses results in SmartLink mode
FTTA-SLM option
Principle of FTTA-SLM
FTTA-SLM is an OTDR software application that is delivered as an option of the OTDR
module (see references in the ordering information section), and is installed onto the
events detected on
trace with shortest pulse
User Manual
Chapter 3 Reflectometry measurements
events detected on
standard trace
78000010220 Rev. 024
FTTA-SLM option
125