Chapter 9 OSCA-710 Optical Spectrum Measurement
Trace display functions
Shift function
The Shift function is used to displace the displayed section of the trace by pressing the
direction keys.
The horizontal shift is performed maintaining the point of intersection between the trace
and the selected cursor at the same level, scrolling the trace horizontally while following
it vertically, so that it never goes off the screen.
To use this function:
1
Select the zoom factor as described above.
2
Choose cursor and cursor position.
3
On the
4
Use the direction keys or touch and hold screen to shift the trace in the desired
direction.
Trace /Table key
This key offers a choice from the following displays:
Trace alone: main display of the trace with a single line of the table at the foot of the
page (see
Figure 162 on page
Trace + Table: display of trace, reduced in size but followed by 5 to 8 lines of the table
of results.
Figure 163
OSCA Results - Trace and Table
322
Zoom/Shift
key, select Shift.
319).
User Manual
78000010220 Rev. 024