Alarms parameters
In the Setup page, press the Alarm menu key (if one parameter is selected in the current
screen, press Top Menu soft key to display the right menu keys and click on Alarms).
Threshold
None
The alarm function is not active.
User
Define your own thresholds values for one or several elements:
Splice Loss / Connector Loss / Reflectance / Fiber Length Min / Link
Loss Max / ORL
TIA-568 C / ISO/IEC 11801 / Default
Select one of this parameter to configure the alarm thresholds with
predefined values:
Splice Loss
Connector Loss
1
Slope
Reflectance
ORL
1. This parameter is not available in OEO-OTDR configuration
If results are above the thresholds, they will be highlighted in red in the table of results,
and the icon
will appear at the top right of the screen.
If all the results lie within the thresholds (no result is in red), results are displayed in
green in the table and the icon is
Analysis
In the Setup page, press Analysis softkey (if one parameter is selected in the current
screen, press Top Menu soft key to display the right menu keys and click on Analysis).
This screen allows to configure the network:
BBU Jumper
Test from the patchcord that will be plugged into BBU.
(not applicable for DAS networks)
Default
> 0.20 dB
> 0.50 dB
> 1.00 dB/km
> - 35 dB
< 27 dB
.
User Manual
Chapter 3 Reflectometry measurements
TIA-568C & ISO/IEC 11801
> 0.30 dB
> 0.75 dB
> 1.00 dB/km
-
-
78000010220 Rev. 024
FTTA-SLM option