Chapter 10 Polarization Mode Dispersion Measurement
Performing a High Resolution PMD measurement
1
To perform a reference, connect your Broadband source 81BBS1A to the PSM
module as shown below:
Figure 171
Interconnection of 81BBS1A and PSM modules
81BBS1A in Base-Unit
2
Press the R
module self-calibration. Wait for completion before continuing
3
On the BBS Results tab, select Source On, enter safety password 4877, and
select PMD/AP function using the soft key.
Figure 172
BBS activation
4
Go to the PMD Setup page, and press the Acq. Ref. softkey.
NOTE
If the 81DISPAP or 81MRDISPAP module with the HR function is used for
the first time, pressing the S
page.
5
Set the parameter
344
Signal input
connector on
the PSM
module
hardkey. A bar graph informs of the progression status of the
ESULTS
Make Zero
User Manual
Coupler
Jumper
PSM module in Base-Unit
button directly open the Reference Setup
ETUP
to
Yes
in the configuration menu.
78000010220 Rev. 024