Perform The Cd/Ap Referencing On T-Berd/Mts 8000 V2; Ap And Cd Referencing With Fc1 Test Kit - Viavi 8100 V2 Series User Manual

Portable, modular platform designed for the construction, validation and maintenance of optical fiber networks
Hide thumbs Also See for 8100 V2 Series:
Table of Contents

Advertisement

1
On the Home page, select BBS icon
2
Press R
ESULTS
3
Click on BBS tab.
4
Click on menu key
5
Press S
TART
Perform the CD/AP referencing on T-BERD/MTS
8000 V2
1
Click on CD or AP tab and press S
2
Click on Ref Acq.
3
Select Make Reference to Yes.
4
Press Yes to confirm making a new reference..
5
On the parameter BBS Serial Number, enter the BBS module serial number
using the numerical keyboard and validate.
NOTE
The serial number is displayed in the About page of the Platform (H
About).
6
Press the soft key
7
Press S
TART

AP and CD referencing with FC1 Test Kit

The referencing for AP and CD in unidirectional mode is performed with:
one T-BERD/MTS 8000 V2 at one side
one OBS5x0 at the other side.
1
Connect a fiber jumper to the MTAU COM port of T-BERD/MTS 8000 V2 and to
the OBS5x0.
2
Connect both jumpers together using the appropriate mating adapter.
See below the interconnection schematic.
hard key.
to select CD or AP.
/S
hard key to activate the BBS.
TOP
to enter in the reference menu.
to exit the reference menu.
/S
hard key to reference
TOP
User Manual
Chapter 17 Fiber Characterization
Referencing the test functions
.
.
ETUP
78000010220 Rev. 024
>
OME
467

Advertisement

Table of Contents
loading

Table of Contents