Chapter 17 Fiber Characterization
Fiber Characterization Automated Test Sequence using the Link Characterization Script
Starting a test sequence
1
Press the soft key Start on both units
2
When message Do you want to start measurement from your site?
appears, click Yes on the 8000 and No on the 6000A
3
Inspect and clean connectors of the fiber to test using the P5000i inspection
scope on both ends when the inspection page appears
4
Connect the fiber under test to each patchcord (one connected to the TB8000
COM port of the MTAU module and one connected the 6000A module C)
5
Press the S
On the T-BERD/MTS 8000 V2
6
Press Yes to the message
between both test sets is displayed.
IL, ORL and bidirectional OTDR measurements are performed under FCOMP
On the T-BERD/MTS 6000A V2
7
The following message appears while the 8000 V2 test is in progress:
DO NOT click until the 8000 V2 notifies test is completed.
On the T-BERD/MTS 8000 V2
1
Notify 6000A user when the prompt
press Yes until far end user confirmed PMD source function is activated
On the T-BERD/MTS 6000A V2
2
When notified by the 8000 V2user, press OK button to continue
3
Touch the softkey
4
Notify the far end user
On the T-BERD/MTS 8000 V2
5
Press Yes to the message
6
Notify 6000A user when the prompt appears
press Yes until far end user confirmed AP source function is activated
492
hard key on both units to continue with the test sequence
CRIPT
to position the test mode as PMD
User Manual
to start PMD measurement
78000010220 Rev. 024
when the green link
appears but DO NOT
but DO NOT