Viavi 8100 V2 Series User Manual page 600

Portable, modular platform designed for the construction, validation and maintenance of optical fiber networks
Hide thumbs Also See for 8100 V2 Series:
Table of Contents

Advertisement

Chapter 21 Technical specifications
High Resolution Dispersion test solution
Measurement time
1. With Broadband Source module E81BBS1A in Corresponding mode
2. With Kit 1 (PMD only). Subtract 6 dB for Kit 2 (PMD / CD / AP)
3. With strong mode coupling
4. Strong mode coupling. 1500-1600nm measurement span. PMD < 10ps
5. Up to 35 dB attenuation
6. Without averaging
Dynamic Range
Wavelength Uncertainty
Measurement Time
Measurement uncertainty
- at 1550 nm
- at 1625 nm
1. With Broadband Source module E81BBS1A in Corresponding mode
2. With averaging
3. Without averaging
4. Measured with 80 km G.652 fiber
Wavelength Acquisition Range
Wavelength Uncertainty
Minimum length
Dynamic Range
Measurement Time
Zero Dispersion wavelength uncertainty
Zero Dispersion wavelength repeatability
Dispersion uncertainty
Dispersion repeatability
1. With Broadband Source module E81BBS1A in Corresponding mode
2. Dynamic range obtained without PSM module. A typical extra budget loss of 1.25dB applies when passing
through PSM
3. Repeatability refers to the typical one sigma standard deviation value, obtained for systems cycling of 20
measurements
4. 1530−1570 nm band
564
Polarization Mode Dispersion
6
Attenuation profile (with kit 2)
1,2
3
4
:
Chromatic Dispersion (with kit 2)
1,2
3
4, 5
3, 4
User Manual
From 30 s.
+/- 0.1 nm
3 seconds
+/- 0.003 dB/km
+/- 0.004 dB/km
1460 - 1640 nm
10s. to 30s.
80km G652
n/a
n/a
+/- 0.06 ps/nm.km
0.02 ps/nm.km
78000010220 Rev. 024
45 dB
+/- 0.1 nm
1 km
33 dB
10km G655
+/- 4.5 nm
0.4 nm
+/- 0.3 ps/nm.km

Advertisement

Table of Contents
loading

Table of Contents