Chapter 9 OSCA-710 Optical Spectrum Measurement
OSCA optical spectrum analyzer Setup

Analysis parameters

In the Setup page, press Analysis softkey (if one parameter is selected in the current
screen, press Top Menu soft key to display the right menu keys and click on Analysis).
Those parameters are only linked to the current active fiber.

Channel Detection

Signal threshold
Threshold of detection of channels (see
Auto.
Manual
Use direction keys or
NOTE
Modification of the parameters
will only modify the results if the OSCA module present is the one that was
used for the acquisition.
Mode
Grid
Permanent
1.Attention: all modification of these parameters has immediate repercussions on the trace
and entails the loss of the measurement statistics.
312
1
1
the threshold is determined automatically.
the threshold can be set from -79.9 to +30 dBm
Edit Number
The grid serves as a detection reference: it must therefore be
Regular, Manual, ITU DWDM, ITU CWDM, LR4/ER4-100G,
LR4/ER4-40G or 10x10-100G. The choice of grid takes priority
over the choice Channel Selection. For example, it is not
possible to choose Channel selection = Grid, if the option
selected for the grid is «Without» or «Conventional».
Automatic detection of the channel on each acquisition. In this
mode the channels are always detected without making a
reference measurement.
User Manual
"Channel detection threshold" on page
to modify values.
Channel detection
78000010220 Rev. 024
and
Signal threshold
323).