Performing A Cd Measurement Through Amplifiers - Viavi 8100 V2 Series User Manual

Portable, modular platform designed for the construction, validation and maintenance of optical fiber networks
Hide thumbs Also See for 8100 V2 Series:
Table of Contents

Advertisement

Figure 214
Connections for the CD Measurement
NOTE
You can modify the fiber length without performing a measurement again.
The CD results (dB/km) will be recalculated automatically.

Performing a CD measurement through amplifiers

The CD reference step is not required for measuring through amplifiers:
1
Connect the source and the 81DISPAP/81MRDISPAP module at each end of the
fiber under test. Optical attenuators (OLA55 1 and OLA55 2) have to be posi-
tioned before and after the amplified link.
Viavi recommends to:
setup the front attenuator (OLA55 #1) so that the power level received by the
amplifier is 10 dB lower than the in-service power level (in normal operation).
setup the far end attenuator (OLA55 #2) so that the maximum received power
level by the CD tester is -10dBm max
Chapter 14 CD measurements using phase shift method
Fiber patchcord
OBS-5XX
User Manual
Performing a CD measurement through amplifiers
Fiber under test
Signal input connector of
the CD module
78000010220 Rev. 024
439

Advertisement

Table of Contents
loading

Table of Contents