Press OK to perform zero ORL referencing.
9
Connect both jumpers together using the appropriate mating adapter. See below
the interconnection schematic.
Figure 232
IL Side by Side Referencing schematic
10
Press Ok to perform IL Referencing.

Perform ORL and Loopback IL Referencing

1
Touch the FCOMP tab.
2
Press the soft key References to enter the reference menu.
3
Press the soft key Take Refs in order to enter the referencing wizard.
4
Press Loopback to start referencing process for ORL and Insertion Loss
5
Repeat above steps 5 to 8 for loopback ORL/IL referencing.

AP and CD referencing with FC2 Test Kit

The referencing for AP and CD in bi-directional mode is performed with:
one T-BERD/MTS 8000 V2 at one side
one T-BERD/MTS-6000A V2, with the E81x6C-FCHAR module and integrated
broadband source, at the other side.
User Manual
Chapter 17 Fiber Characterization
Referencing the test functions
78000010220 Rev. 024