Alarms Parameters - Viavi 8100 V2 Series User Manual

Portable, modular platform designed for the construction, validation and maintenance of optical fiber networks
Hide thumbs Also See for 8100 V2 Series:
Table of Contents

Advertisement

Length
Distance Unit
Define the unit of the distances displayed: km, kfeet, miles, meter, feet
Config.
This parameter displays the configuration file selected for acquisition, and cannot be
modified from Setup page. To modify the configuration file to be used:
1
Click on the menu header FTTH (PON) Setup
2
Click on Load Config. menu key
3
Select the file in the Explorer.

Alarms parameters

In the Setup page, press Alarms softkey (if one parameter is selected in the current
screen, press Top Menu soft key to display the right menu keys and click on Alarms).
None
User
Default / TIA-568 C / ISO/IEC 11801 / G.697/G.98x PON / G.697/IEEE
PON
Select one of this parameter to configure the alarm thresholds with predefined values:
Splice Loss
Connector Loss
Slope
Reflectance
Use the
Edit Number
164.042 kfeet / 31.075 miles) or affect the active cursor value, using
the
Set Cursor Distance
The alarm function is not active.
Define your own thresholds values for one or several elements:
Splice Loss / Connector Loss / Reflectance / Splitter Alarm / Fiber
Length Min / Link Loss Max / ORL
G.697/G.98x PON & G.697/
Default
IEEE PON
> 0.20 dB
> 0.30 dB
> 0.50 dB
> 0.50 dB
> 1.00 dB/km
-
> - 35 dB
> - 35 dB
User Manual
Chapter 3 Reflectometry measurements
key to enter a distance (Min= 0 / Max=50 km /
key.
78000010220 Rev. 024
FTTH-SLM Software option
TIA-568C & ISO/
IEC 11801
> 0.30 dB
> 0.75 dB
> 1.00 dB/km