Chapter 12 High Resolution OSA Measurement

Cleaning of Connectors

Cleaning of Connectors
Cleaning of the patchcord connector is extremely important to avoid damage to the input
of the OSA. A dirty connection can irreversibly damage both surfaces, especially when
working with high powers.
We highly recommend inspection of the patchcord before connecting it to the OSA.

Configuration of the instrument

The instrument configuration menu will be displayed directly after power up or by
pressing the H
The actual module status will be shown (O
1
For configuring the OSA press the OSA-icon (by touchscreen), or select the func-
tion icon by arrow-keys and press E
2
Press R
If the MTS / T-BERD 8000 V2/6000(A) is switched Off in this configuration, the
next start up will directly start the to OSA-application, and display the result
window.
For more details about the general MTS-configuration see the Base Unit manual.
For measurement, connect the fiber to be tested on the optical input of the selected
module.
Kind of input connector:
if the protection cap is green the optical input interface is a angled physical
connector (APC-type)
if it is a black protection cap, the interface type is physical connector (PC-type)

Loading a configuration file

To load the configuration file to be used for HR-OSA test:
1
Press S
2
On bottom right of setup page, press Load Config. menu key.
3
In the Explorer, select the desired file configuration
4
Press
382
-button.
OME
button to see the OSA-result window.
ESULT
hard key.
ETUP
Load
Config. menu keys.
User Manual
/O
) on system window
N
FF
.
NTER
78000010220 Rev. 024