Averaging acquisition
Select between "No" and 64 times averaging.
Resolution
Select the resolution bandwidth used for HR OSA acquisition (depends of the module
used).
Coarse Scan
The coarse mode provides a faster scan over multiple channels at the cost of a
degraded frequency accuracy. Select Yes to perform this scan.
Number of Scans
In Statistics mode, this must be selected between 2 and 1000.
Long Term
NOTE
A long term measurement can be done only if the Sweep parameter is on
Statistics mode.
Long term time diagram:
number of sweeps 7
wait period 5s
1
2
3
4
5
|-----|-----|-----|------|----|-----|
time distance between the next acquisition = wait period (time)
No
the measurement is not automatically launched
Yes
the measurement is automatically done, after the time interval
selected (see hereunder).
6
7
User Manual
Chapter 12 High Resolution OSA Measurement
Configuration of the instrument
78000010220 Rev. 024
385