Operator's check for switching Noise Source
The following procedure checks that the USB Noise Source can be switched
between different modes - Off/On/Normal.
1 Connect the USB Noise Source to the Signal Analyzers using the 8121-3431
Figure 4-4
2 At Meas Setup menu, press the ENR menu key and select NS Setup. Press the
Keysight U1832A/B/C and U1833A/B/C/D User's Guide
USB cable, as shown in
USB Noise Source connection to the Signal Analyzer
Noise Source Preference menu key and set it to SNS (Auto).
Figure
4-4.