Start/Stop the LCA SCPI Module

Keysight N437x Series Lightwave Component Analyzer, Programmer's Guide
All LCA modules require the network analyzer application. It should always
start after a system reboot automatically. If the network analyzer is not
running, please start it manually.
Like all other LCA modules, the LCA SCPI module does not start
automatically. You have to start it manually. Use the
Agilent.LCA.SCPI.Conf.exe program to start or stop the LCA SCPI
interface. To launch this program, you may either use the LCA SCPI link in
the network analyzer GUI macro list under utilities, or the shortcut LCA
SCPI Interface on the desktop or in the program menu.
The LCA SCPI interface is implemented on the LCA Remote Interface
methods and the Agilent Translator Framework. Therefore the LCA Server
starts automatically when you start the LCA SCPI interface. When you click
on the "Start SCPI" button on the SCPI configuration form, the Agilent
Translator Framework starts and loads the Agilent.LCA.SCPI.Module. The
LCA Server cannot handle more than one session. Therefor you can run
either the LCA Measurement Setup application or the LCA SCPI
interface, but not both in parallel. This is the same for the LCA Remote
Client. It also connects to the LCA Server and therefore the SCPI interface
cannot run at the same time.
Remote Operation
1
41