N432A Remote Operation

Setting the Bridge Resistance

A bridge resistance can be configured for the thermistor sensor to balance
bridges.
The [SENSe[1]:]BRESistance <character_data> command allows you to
specify a bridge resistance value of either 100
To check the current setting of the bridge resistance, use the
[SENSe[1]:]BRESistance? query.
For details, refer to

Setting the Measurement Frequency

You can set the frequency of the RF signal you are measuring by using the
[SENSe[1]:]FREQuency[:CW|:FIXed] <numeric_value> command. Setting
this value optimizes the accuracy and minimizes measurement uncertainty,
especially when making comparative measurements between signals.
To check the current frequency setting, use the
[SENSe[1]:]FREQuency[:CW|:FIXed]? [MIN|MAX] query.
For details, refer to
page 377.