Replacement Procedure
Removable Hard Disk Drive Replacement
Step 6. Calibrate the touch screen as described in
WARNING
From the volume label CN925, user recovery function is not prepared. It is required
to inform this change to user before replacing the hard disk drive.

Restore Backup Files onto the New Removable Hard Disk Drive

The backup files are stored into the PC. The file must be restored in the new removable
hard disk drive with the adjustment program. For the details of the adjustment program,
refer to
Step 1. Run the adjustment program.
Step 2. Perform the adjustment using "Recovery" in

Adjustment for the New Removable Hard Disk Drive

Step 1. Run the adjustment program.
Step 2. Perform the adjustment using "HDD" in

Calibration of the Touch Screen

When you have replaced the HDD on a E5052B equipped, you have to calibrate the touch
screen. Follow the procedure described below to calibrate the touch screen.
Step 1. Press
Step 2. Press
Step 3. Press
Step 4. Press
The touch screen calibration screen
144
Chapter 6 on page
133.
.
.
Service Menu
Test Menu
.
.
Adjust Touch Screen
"Calibration of the Touch Screen" on page
Spot Adjustment
of the program.
Spot Adjustment
(Figure
6-5) appears.
144.
of the program.
Chapter 6