Replacement Procedure

Removable Hard Disk Drive Replacement (Obsolete)
Replacement Procedure
Step 1. Replace the removable hard disk drive by inverse procedure of removal.
Fasten the TORX T10 screws (item 6) using a torque screwdriver.
Step 2. Insert the hard disk into the slot.
Step 3. Turn the key to the left until it is locked, then remove the key.
The "U" is displayed at the 8 segment LED on next to the removal hard disk
when E5071C is turned on without key lock.
Step 4. Restore backup files onto the new removable hard disk drive as described in
"Restore Backup Files onto the New Hard Disk Drive" on page
Step 5. Calibrate the touch screen as described in
on page

Restore Backup Files onto the New Hard Disk Drive

The following procedure shows how to write the system correction data stored
in the backup memory to the newly mounted hard disk.
Step 1. Turn on the E5071C.
Step 2. Press
Step 3. Click Load Project... in the softkey menu.
Step 4. Select RestoreSysCorFile.vba from the D:\Agilent\Service folder and then
press the Open button.
Step 5. Press
Step 6. The RestoreSysCorFile dialog box appears.
Step 7. Click OK.
Once the program has terminated, restart the E5071C to enable the
system correction data.