Reading/Writing Correction Data - Keysight E4980A User Manual

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

Configuring Measurement Conditions (Display and Function Related Settings)
CORRECTION page

Reading/Writing Correction Data

The E4980A/AL allows you to read and write the correction data.
To write the correction data:
Step 1. Press the [Meas Setup] key.
Step 2. Press the CORRECTION softkey.
Step 3. Using the cursor keys, select the SPOT No. field and enter the measurement
point number.
Step 4. Using the cursor keys, select the FREQ field and enter the frequency.
Step 5. Using the cursor keys, select these fields: OPEN A, SHORT A, LOAD A, OPEN
B, SHORT B, and LOAD B.
Step 6. Enter the correction data using the softkeys or entry keys. If you use the entry
keys to enter the value, the softkey labels change to unit labels (p, n, u, m,
x1).
126
Keysight E4980A/AL Precision LCR Meter