4
Remote Programming
Calibration Commands
:CALibrate:ADC[:STARt]
:CALibrate:ADC:ABORt
:CALibrate:ADC:INFormation
Returned Format
Example
90
The CALibration subsystem commands allow to re-calibrate the ADC of
the digitizer, to query the calibration status and to save and load
calibration data.
Command
:CALibrate:ADC[:STARt]
Description
The command starts the self-calibration of the ADC module.
Command
:CALibrate:ADC:ABORt
Description
The command aborts a running self-calibration of the ADC module.
Query
:CALibrate:ADC:INFormation?
Description
The query returns information about the last calibration of the ADC
module.
[:CALibrate:ADC:INFormation] <information_string><NL>
<information_string> Contains information about type ("User" or
"Factory") and date of last calibration.
Query
:CALibrate:ADC:INFormation?
"M8131A Setup:
Mode: 16GHz on 4 channels
Calibration Loaded:
1(User): 2018-03-09 17:33:44"
Keysight M8131A 16/32 GSa/s Digitizer User's Guide