Probe Calibration

To Calibrate the Probe on Infiniium Oscilloscopes

To Calibrate the Probe on Infiniium Oscilloscopes
Procedure
66
During this calibration, you may be prompted to make one or more
adjustments. These adjustments, located on the underside of the
probe amplifier, are shown in
Table 8
Test Equipment
N2825A User-Defined
Probe Head
Pomona 3788 BNC
Female to Minigrabber
Test Clips Breakout
Adapters
Screwdriver
Redirecting the Probe Comp Signal
NOTE
On 9000-series and S-series Infiniium oscilloscopes, you can redirect
the Probe Comp signal to the Aux Output BNC connector on the rear
panel of the oscilloscope. This connector can then be connected to the
Pomona 3788 using a BNC cable. Click Utilities > Calibration and select
Probe Comp in the Aux Output drop-down list.
1
Turn the oscilloscope on.
2
Solder the sensor leads onto the N2825A user-defined head as
explained in
3
Attach the N2825A user-defined head to the probe.
Figure 32
Required Calibration Test Equipment
Description
Provided with probe.
SMA, BNC, BNC Tee, BNC cables and cable clips as needed.
For making adjustments during the procedure. This tool is
provided with the probe.
"Quick Start"
on page 23.
on page 69.
Refer to
http://www.pomonaelectronics.com/
for more information and how to order
this product.
N2820/1A User's Guide