Zero/Cal Lockout; Figure 2-42 Please Zero And Calibrate Window - Keysight N1913 User Manual

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

General Power Meter Functions
2

Zero/Cal Lockout

The Zero/Cal Lockout facility provides a mean of ensuring that a measurement
cannot be taken until the connected sensor has been zeroed and calibrated. If the
Zero/Cal Lockout facility is enabled and a sensor is connected which have not
been zeroed and calibrated, then the display window for the sensor will display
the message Please Zero and Cal.
Figure 2-42
Please zero and calibrate window
When you zero the sensor, the message changes to Please Cal. If you calibrate
the sensor before zeroing it, the message changes to Please Zero.
Dual channel meters display channel specific messages when a sensor is
connected. The Zero/Cal Lockout configuration is applied to both channels - it
cannot be applied to one channel only.
You can enable and disable the Zero/Cal Lockout facility from the System menu or
the Cal menu as follow:
Press
,
,
or
. Similarly, press
,
,
1 of 2
Must Cal Off
On
1 of 2
or
.
Must Cal Off
On
Keysight N1913/1914A User's Guide
103