Ap Reference Measurement; Performing The Reference - Viavi ODM V2 User Manual

Optical dispersion measurement module
Table of Contents

Advertisement

Chapter 3 Attenuation profile

AP Reference Measurement

AP Reference Measurement
On the measurement examples below, only the OBS-5XX is described. However, it can
be replaced by E81BBS2A.

Performing the reference

It is mandatory to perform a broadband source referencing once a day or each time the
fiber patchcord has to be changed.
To make a reference, connect your OBS-5XX or BBS to the test module just like shown
on figure below.
Figure 24
Reference measurement setup witn an OBS-5XX
OBS-5XX
1
Activate the source on the distant OBS and select AP function.
If a BBS Module is used, press Source START
password 4877 to activate the Source.
40
Mating adapter
Fiber patchcord
User Manual
700MAN902 Rev. 000
and enter the