Function Overview
Aperture time is the time required to acquire the measurement data. For accurate
and reliable measurement, the aperture should be increased. This value can be
specified.
Overhead time includes other factors such as range change or data
compensation. This value depends on the measurement conditions and cannot be
specified.
Remote Transient Voltage Measurement
The M9602A/M9603A has a dedicated voltmeter with higher bandwidth for the
remote transient voltage measurement. This function is effective in measuring
pulse voltage with the width of a few to dozens of microseconds. You can enable
or disable this function.
Automatic Output On/Off
The M9602A/M9603A module has output on/off settings when the trigger system
changes the status. You can enable or disable these settings independently.
Auto Output On
Auto Output Off
Output Off Mode
The output off mode is the condition immediately after the source output is turned
off. The mode should be specified by programming before enabling the source
output. The following settings are available for the output off mode.
High impedance Output switch: off
Normal
42
The source output is automatically turned on just before the
trigger system is initiated by programming.
When the trigger system is initiated on the Soft Front Panel,
this setting is ignored.
The source output is automatically turned off when all trigger
systems change the status from busy to idle.
Voltage source setup is not changed.
The current source setup is not changed if the current output
value and range are less than or equal to 10 mA and 10 mA
range. Otherwise, the output value is set to 10 mA, and the
output range is set to 10 mA range.
Output switch: off
Source function: voltage source
Keysight M9602A/M9603A Startup Guide, Edition 3