Instrument Messages
Condition Messages

Condition Errors 801 to 899, Temperature

This series of errors corresponds to the bits in the
STATus:QUEStionable:TEMPerature register. The second column in the table
shows the corresponding bit.
An event with the error number shown in the table means the condition has
been detected. When the condition is cleared, an event with the error number
plus 1000 is generated. These error numbers can be seen in the Show Errors
screen, along with the DETECTED and CLEARED indicators.
For example, error 801 indicates that the Ref Osc Oven Cold condition has
been detected; error 1801 indicates that condition has been cleared.
This register is summarized as bit 4 of the STATus:QUEStionable register. See
section
Err#
Bit in
status
register
801
0
803
1
805
2
807
3
809
4
811
5
813
6
815
7
817
8
819
9
821
10
823
11
825
12
827
13
829
14
Keysight Technologies N9040B UXA Signal Analyzer Service Guide
"Condition Errors 601 to 699, Error Summaries" on page
Message
Reference Oscillator Oven
Cold
unused
unused
unused
unused
unused
unused
unused
unused
unused
unused
unused
unused
unused
unused
Error or
More Information
Warning
W
(not currently in use)
134.
137