Status:questionable:condition - Keysight U2722A Programmer's Reference Manual

Usb modular source measure units
Hide thumbs Also See for U2722A:
Table of Contents

Advertisement

STATus:QUEStionable:CONDition?

Syntax

STATus:QUEStionable:CONDition?

This query returns the value of the Questionable Status Condition register. The
Condition register is a read-only register, which holds the live (unlatched)
questionable status of the instrument. Reading the Questionable Condition
register does not clear it.
For more information on the SCPI Status System, refer to
NOTE
system diagram"
Remarks
The following table lists the bit definitions for the Questionable Status registers.
Bit number
0 to 3 Not Used
4 OT
5 to 15 Not Used
Returned Query Format
<NR1>
The returned value is the binary-weighted sum of all bits set in the register. For
example, if bit 4 is set, this query command will return "+16".
Example
The following query command reads the Questionable Status Condition register.
STAT:QUES:COND?
Typical Response: +16
Keysight U2722A/U2723A Programmer's Reference
on page 21.
Decimal value
0
16
0
Definition
0 is returned.
The over-temperature protection has tripped.
0 is returned.
STATus Subsystem
Chapter
2,
"Status
10
93