Concluding the setup for all options

NOTE
Keysight N4391A/N4392A Optical Modulation Analyzer, Getting Started
1
At the rear of the instrument,
a Connect the mouse and keyboard to the digital signal analyzer.
b Using the USB cable supplied, connect the optical modulation
analyzer to the digital signal analyzer.
c Using the two power cables supplied, connect the power cables to
the digital signal analyzer (the power socket with the notch), and the
optical coherent analyzer.
d If necessary, connect your LAN cable to the digital signal analyzer.
2
At the front of the instrument, put the electrostatic discharge band
around you wrist, and plug it into the earthed connector at the bottom
right of the instrument.
3
Working from one side to the other (left to right or right to left), attach
the rigid cables between the digital signal analyzer and the optical
modulation analyzer.
Note: you may find this easier if you move the front of the instrument
so it is at the front edge of the table or bench you are working on.
a Remove the protective cap from the signal input on the optical
modulation analyzer.
b Make sure you have the rigid cable with the same number as the
connector on the digital signal analyzer.
c Remove the protective cap from the smaller connector on the rigid
cable.
d Connect the BNC connector to the digital signal analyzer.
Tighten the connector until it is finger tight.
e Connect the rigid cable to the connector on the optical modulation
analyzer.
The connectors on the optical modulation analyzer tighten anti-clockwise.
f
Using the spanner wrench supplied, holding the end of the rigid
cable, and use the torque wrench supplied to tighten the connector.
Getting Started
2
29