Advertisement

SCPI Command Reference
E4980A/AL commands
Parameter
ON or 1
OFF or 0 (Preset value)
Equivalent key
[Display Format] - BIN COUNT - COUNT - COUNT ON
[Display Format] - BIN COUNT - COUNT - COUNT OFF

:COMParator:MODE

Syntax
:COMParator:MODE {ATOLerance|PTOLerance|SEQuence}
:COMParator:MODE?
Description
Selects the limit mode of the comparator function. This command clears the
primary and secondary parameter limit values.
Parameter
ATOLerance
PTOLerance (Preset value)
SEQuence
Equivalent key
[Meas Setup] - LIMIT TABLE - MODE - ABS
[Meas Setup] - LIMIT TABLE - MODE - %
[Meas Setup] - LIMIT TABLE - MODE - SEQ

:COMParator:SEQuence:BIN

Syntax
:COMParator:SEQuence:BIN <Bin 1 Low>,<Bin 1 High>,... ,<Bin n High>
:COMParator:SEQuence:BIN?
Description
Sets the low/high limit values of the BINs for the sequential mode of the
comparator function. These limits can be set only when the limit mode is set in
the sequential mode. Of course, the lower limit value must be lower than that
of the upper limit. If a BIN is not used, this command returns the values of
-9.9E37 and 9.9E37 for the lower and upper limits, respectively.
Keysight E4980A/AL Precision LCR Meter
Description
Enables the BIN count function
Disables the BIN count function
Description
Sets the Comparator mode. This command clears the
limit table data and secondary limits at "ABS."
Sets the Comparator mode. This command clears the
limit table data and secondary limits at "%."
Sets the Comparator mode. This command clears the
limit table data and secondary limits at "SEQ."
327